|
Membrane Cleaning Procedures We do not recommend cleaning your membrane unless you are sure that it is indicated. With a Spectra System cleaning is required when higher than normal pressures are encountered and the water production has fallen. Always check the prefilters first. Every cleaning will shorten the life of the membrane. For normal cleaning, the SC-3 Acid Cleaning Compound is used first, then the SC-2 Alkaline Cleaning Compound. If known bio-fouling is present, the SC-2 may be used first. Using hot water if possible, up to but not to exceed 120º, is highly recommended as it greatly enhances the ability of the cleaners to do their jobs. If the history of the system is unknown or has been left "unpickled" for an extended length of time and biological growth is present, it is recommended that the system is cleaned with the SC-2, using an alternate source of unchlorinated fresh water before the system is run under pressure. A simple test can be performed to see if biological growth has occurred: Before running the system, remove the pre-filters and examine their condition. If the housings are full of discolored smelly water, the system was not properly stored. Install clean pre-filters if they were bad. Next check the membrane. Attach the brine discharge service hose and lead to a container. Open the pressure relief valve 1 turn and turn the brine discharge service valve to the "Service" position. Manually run the system for 30 seconds. Examine the brine water; if its discolored and smells bad, perform a SC-2 cleaning with an alternate source of unchlorinated water before running the system pressurized. If the brine is fairly clean, the system can be purged, run normally, and checked for performance. Clean the membranes only if performance is reduced. Heating the water is preferable. One way to do this is find a camp stove and use a large SS pot to heat the solution in. The cleaning solution throughout the system will heat as it circulates in and out of the pot. An alternative is to heat the one or two gallons of initial water in the cleaning container to up to 120° F on the main stove before mixing in the cleaner and circulating it into the system. Periodically stop and reheat the solution. Perform the cleaning procedures while the ship is in acceptable sea water for purging and testing. Cleaning Procedures Note: procedures are the same for the SC-3 and SC-2 cleaners A Spectra Cleaning Compound (SC-2 or SC-3) must be mixed with fresh water at a ratio of 1 Spectra container to 3 gallons to have the proper solution. An average of two gallons of water is in a 400 system or 4 gallons in a 1000 system. This water has to be figured in the mixture. A 400 system uses one cleaner container, a 1000 system uses two containers. SC-3 and SC-2 are never mixed together. Do not use them for storage. Warning! The pressure relief valve on the Clark pump must be open for this procedure or membrane damage may result. Maximum pressure 50 psi. Note: Triple the run/soak times if the solution is not heated
|

